Hydraulic forklift scales are also known as: trailer scales, truck scales, forklift scales, ground cattle and so on. The forklift scale has both the forklift handling function and the weighing function of the electronic scale, and the multi-function electronic scale is combined into one.
Fault analysis and solution for hydraulic forklift scales:
Fault 1: The forklift scale is weighed many times, the weight is inconsistent, and there is an error.
Solution 1: Lift the forklift scale to zui high, insert a piece of paper into the gap between the scale and the fork cover to see if there is any place. If there is a problem, you can use the appropriate tool to lift off.
Solution 2: Raise the fork cover, loosen the sensor retainer, use a large screwdriver to adjust it up, and then attach the circlip. If there is no adjustment screw on the sensor, loosen the circlip and then remove the fork cover, then turn the 4 screws in the cover down 2-3 times.
Fault 2: When the forklift scale is weighed, the hop count does not return to zero.
Solution: First of all, if you find that the weighing is not accurate, that is, an item is the same under the two weighing conditions, you need to recalibrate the debugging until it is accurate. If the hop count produces non-return to zero error, please check whether the sensor fixing screw is loose, whether it is in a plane. If no screw is loose and in a plane, check the sensor itself for any problem (sensor self-test method: test with universal meter) The four sensors are blocked to see if the linearity is the same. If the sensor itself has a problem, the sensor needs to be replaced.
Fault 3: The forklift scale weighs no response and often reports errors.
Solution: When you encounter this problem, first check if there is any problem with the line, if there is a short circuit at the end of the wire, if the inner wire of the junction box is off, and so on. Secondly, check if the meter can be turned on normally and perform self-test. If the problem is caused by the line, the wiring should be connected. If the instrument itself has problems, please contact the manufacturer/or distributor.
Fault 4: The monitor cannot be turned on normally, power is on, and the calibration procedure continues to report errors.
Solution: First check if the monitor power supply can be supplied normally (whether the battery has power). If everything is normal, please contact the manufacturer and/or distributor.